Implementation of the International System for Reporting Serous Fluid Cytopathology (ISRSFC): Diagnostic Concordance Analysis between Liquid-based Cytology Preparation Methods
Abstract
Objective: To evaluate the diagnostic concordance between EasyPrep and ThinPrep liquid-based cytology systems using the International System for Reporting Serous Fluid Cytopathology (ISRSFC).
Material and Methods: This prospective study analyzed 555 serous effusion specimens (316 pleural, 218 peritoneal, 21 pericardial) processed in parallel using both LBC methods. Two cytopathologists independently categorized cases according to ISRSFC: nondiagnostic (ND), negative for malignancy (NFM), atypia of undetermined significance (AUS), suspicious for malignancy (SFM), and malignant (MAL). Diagnostic concordance, risk of malignancy (ROM), and performance parameters were calculated with histopathological correlation.
Results: Overall diagnostic concordance was 94.4% with a weighted kappa of 0.912 (95% CI: 0.887-0.937). Distribution for ThinPrep vs EasyPrep: ND 2.5% vs 2.9%, NFM 71.0% vs 70.1%, AUS 2.2% vs 2.7%, SFM 2.3% vs 2.0%, MAL 22.0% vs 22.3%. ROM showed no significant differences between methods: NFM 7.8% vs 8.2%, AUS 60.0% vs 58.3%, SFM 81.8% vs 80.0%, MAL 98.3% vs 97.5%. Both methods demonstrated comparable diagnostic performance: sensitivity 89.5% vs 90.1%, specificity 96.8% vs 96.3%, accuracy 93.6% vs 93.7% (all p-value>0.05). Inter-observer agreement was substantial (κ=0.782-0.798). Minor cytomorphological differences did not impact diagnostic interpretation.
Conclusion: EasyPrep and ThinPrep systems show excellent diagnostic concordance for ISRSFC implementation. Both methods provide equivalent ROM assessment and diagnostic accuracy, supporting laboratory flexibility in LBC platform selection while maintaining standardized reporting quality.
Keywords
Full Text:
PDF (Early Version)References
Chandra A, Crothers B, Kurtycz D, Schmitt F. The International System for Serous Fluid Cytopathology. 1st Edition. Cham, Switzerland: Springer International Publishing; 2020. doi: 10.1007/978-3-030-53908-5.
Sachan R, Gupta A, Awasthi PN, et al. Application of international system for reporting serous fluid cytology (ISRSFC) in effusion samples-a prospective study in an oncology setting. J Am Soc Cytopathol. 2023;12:351-61. doi: 10.1016/j.jasc.2023.04.005.
Rodriguez EF, Jones R, Gabrielson M, et al. Application of the International System for Reporting Serous Fluid Cytopathology (ISRSFC) on Reporting Pericardial Effusion Cytology. Acta Cytol 2020;64:477-85. doi: 10.1159/000507311.
Ahuja S, Malviya A. Categorisation of serous effusions using the International System for Reporting Serous Fluid Cytopathology and assessment of risk of malignancy with diagnostic accuracy. Cytopathology 2022;33:176-84. doi: 10.1111/cyt.13089.
Kolte S, Zaheer S, Aden D, Ranga S. Application of the international system for reporting serous fluid cytology on reporting various body fluids; experience of a tertiary care hospital. Cytojournal 2022;19:52. doi: 10.25259/Cytojournal_49_2021.
Im S, Chae S, Lee D, et al. Comparison of diagnostic accuracy between CellprepPlus® and ThinPrep® liquid-based preparations in effusion cytology. Diagn Cytopathol 2014;42:384-90. doi: 10.1002/dc.23051.
Rajeswaran PK, Srinivasan V, Mahesh SSV, Ashok AU. A retrospective analysis of the application of the newly proposed international system for reporting serous fluid cytopathology on serous effusion specimens: an institutional experience. Acta Cytol 2023;67:70-9. doi: 10.1159/000527398.
Pinto D, Chandra A, Crothers BA, Kurtycz DFI, Schmitt F. The international system for reporting serous fluid cytopathology-diagnostic categories and clinical management. J Am Soc Cytopathol. 2020;9:469-77. doi: 10.1016/j.jasc.2020.05.015.
Castilho da Silva DJ, Dos Santos CR, Xavier-Júnior JCC. Risk of malignancy in effusions according to the international system for serous fluid cytopathology: a review. Acta Cytol 2024;68:384-393. doi: 10.1159/000540414.
Yang H, Zhu J, Wang P. Application of the International System for Reporting Serous Fluid Cytopathology (ISRSFC) in reporting serous effusion: a retrospective study. Medicine (Baltimore) 2023;102:e35707. doi: 10.1097/MD.0000000000035707.
Alashetty S, Sadasivan B, Dharmalingam P, Rajagopal N, Kavya L, Pai MM. The Role of novel tiered reporting system in serous fluid cytology and risk of malignancy assessment: a retrospective study in a tertiary care center. J Cytol 2023;40:107-13. doi: 10.4103/joc.joc_107_22.
Wang M, Chandra A, Cai G. The international system for reporting serous fluid cytopathology—an updated review. J Clin Transl Pathol 2023;3:160-77. doi: 10.14218/jctp.2023.00025.
Song MJ, Jo U, Jeong JS, Cho KJ, Gong G, Cho YM, et al. Clinico-cytopathologic analysis of 574 Pericardial Effusion Specimens: Application of the international system for reporting serous fluid cytopathology (ISRSFC) and long-term clinical follow-up. Cancer Cytopathol 2021;129:980-9. doi: 10.1002/cncy.22510.
Tabatabai ZL, Nayar R, Souers RJ, Crothers BA, Davey DD. Performance characteristics of body fluid cytology analysis of 344,380 responses from the college of American pathologists interlaboratory comparison program in nongynecologic cytopathology. Arch Pathol Lab Med 2018;142:53-8. doi: 10.5858/arpa.2016-0509-CP.
Refbacks
- There are currently no refbacks.

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.