Home > Vol 43, No 2 (2025) > Prasanchaimontri

Comparison of the Radiation Protection Effectiveness between Lead and Antimony-Tungsten Bed Skirt in Fluoroscopic-guided Simulated Urological Surgery

Phornphen Prasanchaimontri, Waraporn Sudchai, Peerawan Sangondee

Abstract

Objective: This study aimed to develop a new shielding device, a bed skirt, and compare the radiation protection effectiveness between lead and antimony-tungsten (Sb-W) bed skirts in fluoroscopic-guided urological surgery. 
Material and Methods: The simulated surgery of ureteroscopy (URS) and percutaneous nephrolithotomy (PCNL), using the phantoms, was set. The lead and Sb-W skirts were mounted around the operating table. The fluoroscopy time was 120 seconds, and the experiment was repeated 10 times. The radiation dose to the body parts of the patient, surgeon, and anesthesiologist was measured with optically stimulated luminescent dosimeters and compared in the following conditions: conventional operating table, Sb-W bed skirt, and lead bed skirt.
Results: By installing a bed skirt, the radiation dose was decreased, with lead being dominant in all areas. The lead bed skirt reduced the radiation dose by 20% to the patient, 66% to the surgeon, and 91% to the anesthesiologist. The absorbed radiation dose of the lead bed skirt was lower than the Sb-W bed skirt in URS settings (23.3±1.8 µGy vs 32.0±1.9 µGy, p-value<0.001) as well as PCNL settings (257.2±15.6 µGy vs 296.5±24.6 µGy, p-value<0.001). There was no statistically significant difference between the conventional operating table and the Sb-W bed skirt in both URS settings (p-value=0.066) and PCNL settings (p-value=0.153).
Conclusion: The lead bed skirt significantly reduced the radiation exposure and provided superior radiation protection compared with the Sb-W bed skirt. This shielding technique is practical to minimize the harmful effects of radiation from fluoroscopy.

 Keywords

fluoroscopy; percutaneous nephrolithotomy; radiation exposure; radiation protection; ureteroscopy

 Full Text:

PDF

References

Geraghty RM, Davis NF, Tzelves L, Lombardo R, Yuan C, Thomas K, et al. Best practice in interventional management of urolithiasis: an update from the European association of urology guidelines panel for urolithiasis 2022. Eur Urol Focus 2023;9:199–208.

Vanzant D, Mukhdomi J. StatPearls. Trasure Island: StatPearls; 2023.

Gogos S, Touzell A, Lerner LB. What we know about intra-operative radiation exposure and hazards to operating theatre staff: a systematic review. ANZ J Surg 2022;92:51–6.

Stahl CM, Meisinger QC, Andre MP, Kinney TB, Newton IG. Radiation risk to the fluoroscopy operator and staff. AJR Am J Roentgenol 2016;207:737–44.

Vano E, Kleiman NJ, Duran A, Romano MM, Rehani MM. Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol 2013;24:197–204.

Jacob S, Boveda S, Bar O, Brézin A, Maccia C, Laurier D, et al. Interventional cardiologists and risk of radiation-induced cataract: results of a French multicenter observational study. Int J Cardiol 2013;167:1843–7.

Jacob S, Michel M, Spaulding C, Boveda S, Bar O, Brézin AP, et al. Occupational cataracts and lens opacities in interventional cardiology (O’CLOC study): are X-rays involved? radiation-induced cataracts and lens opacities. BMC Public Health 2010;10:537.

Bhanot R, Hameed ZBM, Shah M, Juliebø-Jones P, Skolarikos A, Somani B. ALARA in urology: steps to minimise radiation exposure during all parts of the endourological journey. Curr Urol Rep 2022;23:255–9.

Thatcher J, Mueller T, Klimowich K. Don’t skirt the question: lead bed skirts in the operating room, an intervention. J Urol 2021;206:e130-1.

Giblin JG, Rubenstein J, Taylor A, Pahira J. Radiation risk to the urologist during endourologic procedures, and a new shield that reduces exposure. Urology 1996;48:624–7.

Yang RM, Morgan T, Bellman GC. Radiation protection during percutaneous nephrolithotomy: a new urologic surgery radiation shield. J Endourol 2002;16:727–31.

Amirhasani S, Daneshdoost R, Mousavibahar S, Ghazikhanlou-Sani K, Raeisi R. Reduction of radiation dose received by surgeons and patients during percutaneous nephrolithotomy: a new shielding method. Urol J 2021;18:271–6.

Kumari G, Kumar P, Wadhwa P, Aron M, Gupta NP, Dogra PN. Radiation exposure to the patient and operating room personnel during percutaneous nephrolithotomy. Int Urol Nephrol 2006;38:207–10.

Burns KM, Shoag JM, Kahlon SS, Parsons PJ, Bijur PE, Taragin BH, et al. Lead aprons are a lead exposure hazard. J Am Coll Radiol 2017;14:641–7.

Manocchio F, Ni T, Pron G, Jaffer H, Murphy K. Lead-dust contamination on radiation protection apparel. J Vasc Interv Radiol 2023;34:563–7.

Sweilum O, El-Badry A, El-Agamy S. Risk assessment of lead exposure in radiology personnel in Menoufia hospitals, Egypt. Toxicol Ind Health 2023;39:281–9.

Lu H, Boyd C, Dawson J. Lightweight lead aprons: the emperor’s new clothes in the Angiography Suite? Eur J Vasc Endovasc Surg 2019;57:730–9.

König AM, Verbe Zoum J, Fiebich M, Abissi PW, Mahnken AH. Comparison of the radiation protection effect of different radiation protection aprons made of different materials. Eur J Radiol 2023;164:110862.

DOI: http://dx.doi.org/10.31584/jhsmr.20241117

Refbacks

  • There are currently no refbacks.
Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

SUBMIT A PAPER

JHSMR accepts online submission through

AHR-iCON 2024

Journal Metrics


2020
Acceptance rate: 52%
2021
Acceptance rate: 27.8%
2022 (March)
Acceptance rate: 15.6%
2023 (June)
Acceptance rate: 23.6%


Submission to final decision
74 days

Acceptance to publication
40 days

SCImago Journal & Country Rank

About The Authors

Phornphen Prasanchaimontri
Department of Surgery, Ratchaburi Hospital, Mueang, Ratchaburi 70000,
Thailand

Waraporn Sudchai
Nuclear Technology Service Center, Thailand Institute of Nuclear Technology, Ongkharak, Nakhon Nayok 26120,
Thailand

Peerawan Sangondee
Nuclear Technology Service Center, Thailand Institute of Nuclear Technology, Ongkharak, Nakhon Nayok 26120,
Thailand

Article Tools
Abstract
Print this article
Indexing metadata
How to cite item
Email this article (Login required)
Email the author (Login required)

Supported by

 

JHSMR now Indexed in



Scopus logo.svg






Image result for crossref





PSUMJ Homepage

Keywords COVID-19 SARS-CoV-2 Thailand children computed tomography depression diabetes diabetes mellitus elderly factors hypertension knowledge mental health mortality prevalence quality of life reliability risk factors stroke treatment validity
Journal Content

Browse
  • By Issue
  • By Author
  • By Title
Font Size
Make font size smaller Make font size default Make font size larger

Open Journal Systems